AIR SAMPLING INSTRUEMTS |
|
| |
|
|
| |
| Specifications: |
| Flow Rate |
0.9-1.4m3/Minute. |
Particle size |
10 to 0.5 Micron on filter
paper. More than 10 Micron should be collected in a separate
sampling bottle. |
| Sampling time |
24 hrs (Maximum) |
| Sampling time record |
99.59 hrs. |
| Automatic Sampling Control |
Automatic shut of system
after preset time interval |
| Power requirement |
270V to 230V Single phase
50 HZ. to 60 Hz. built in voltage stabilizer with automatic
shut off beyond 170-270V. |
|
|
| |
|
|
| |
| |
Special
Features:
-
High capacity Blower Motor (Rallifowlf
Make) of 2m3/min.
-
Blower Motor Protection circuits against
under/over voltage and excess current situations.
-
Flow measurement with U tube
manometer and orifice plate at exhause as per US EPA
Guideline.
-
Electronic Timer cum time totaliser
with quartz clock accuracy and automatic start after set
delay time.
-
Gaseous sampling impingers should be
double walled cold housing.
|
|
|
| |
|
|
| |
STACK MONITORING KIT |
|
| |
|
|
| |
| Specifications: |
| Stack Temperature |
Ambient to 600 deg C displayed on a digital
pyrometer. |
| Themocouple |
Theromocouple sensor should
be in stainless steel casing length of insertions should
be one meter with atleast two meter long cable. |
| Stack velocity range and Manometer |
0-30 meter per second should
be able to measure by glass borosilicate tube attached with
spirit label for labelling and for measuring of low velocity
differential density manometer should be provided. |
| Rotameters |
Two rotameters with 2 percent
FSD accuracy of 0-30 1pm and 0-3 1pm should be provided
for particulate and gaseous sampling. |
| Pitot Tube |
S type Pitot tube fabricated
from stainless steel should be of one meter length. |
| Filter Holder |
Should be fabricated from
steel and should be suitable to hold both cellulose thimble
of 30x100 mm and GF/A thimble of 19x19 mm. |
| Nozzles |
A set of 3 nozzles of stainless
steel should be provided. |
| Digital clock |
A stop watch of 0-60 minute
with 1 seconfd read out should be provided. |
| Cold box |
A cold box should be provided
with a provision to keep ice and three impinger tubes with
all-socket joints. |
| Vaccum pump |
Rotary. Vane type free
air displacement 50 LPM driven by single phase motor. |
|
|
| |
|
|
| |
|
|
| |
|
|
| |
AMBIENT FINE DUST SAMPLER |
|
| |
Model
PEM-ADS 2.5u /10u
Features:
- PM 2.5 sampler as per US EPA specifications
- PM 10 or TSP monitoring options
- Full data logging
- Electronics Display of al parameters
- Filter sample allows further analysis
- Sturdy construction for outdoor use
- Simple to set up
- Pocket sized data logger module for data recovery (optional)
- Software calculates final result from user data input (optional)
|
|
| |
|
|
| |
Specifications
The sampler consists of sampler air inlet (designed as per EPA guidelines),
down tube, 2.5u particle size separator (based on EPA WINS impactor),
filter holder assemble, air pump & flow rate control system, flow
rate measuring system based on orifice metering, ambient & filter
temperature monitoring system, barometric pressure measurement sensor,
electronic timer (Real time), out door environmental enclosure &
electronic control system to meet the design & functional performance
as specified by US EPA.
|
|
| |
|
|
| |
| PM 2.5 Size Separator |
: |
EPA Omni-directional ambient particle inlet with sample
transport tube & EPA PM 2.5 WINS Impactor. |
| |
|
|
| Filter Media |
: |
47 mm PTFE Filter in a suitable filter Holder assembly. |
| |
|
|
| Electronic Unit |
: |
|
- Flow Rate Meter & Volume Totaliser
|
: |
Flow in LPM on digital display with a resolution of 0.1 LPM
under actual
operating conditions.
Range 16.7-LPM ± 10% Accuracy ± 2% of reading.
Volume Totalised every 30 seconds from above flow rate &
available on
display. |
| |
|
|
- Flow Control System
|
: |
Electronic sample air flow rate control system based on Flow
Control Proportioning valve to operate over a pressure drop
of clean filter + 60 mm Hg. Flow regulation better than ±
5% over 24 hour period at a rate of 16.7 LPM. |
| |
|
|
- Clock / Timer system
|
: |
Programmable Real Time Clock timing / control system with
digital display of date & time of sampling. Accuracy ±
2 min / month. |
| |
|
|
- Temperature Sensors
|
: |
For Ambient temperature & Filter temperature Range –
5 °C to 50°C with a Resolution of 0.1°C. |
| |
|
|
- Barometric Pressure Sensor
|
: |
Range 600 to 800 mm Hg. Resolution 1 mm Hg. |
| |
|
|
- RS-232 Serial Output
|
: |
For downloading to data storage Module. |
| |
|
|
| Vacuum Pump |
: |
A diaphragm type pump with AC motor. |
| |
|
|
| Power |
: |
230 V AC ± 10%, 50 Hz. (110 V, 60 Hz Optional) |
| |
|
|
| Height of Sampling Inlet |
: |
2 Meter above ground. |
|
|
| |
|
|
| |
|
|
| |
|
|
| |
High Volume Samplers |
|
| |
Model
PEM – HVF 3 & 310
Features:
- Microprocessor based design
- Digital display of Sampling Flow rate and Sampling Time
- Digital display of STP corrected Volume of Sampled air.
- Solid State pressure sensor instead of U-tube manometer for
flow measurement.
- Real time programmable for automatic Start & Stop.
- Brushless maintenance free high capacity Blower.
- Data stored in non-volatile memory.
- Double Walled Cold Housing for 4 Impinger Tubes.
- Blower Motor Protection against Under – Voltage / Over-Voltage
as well as Over Heating of the Motor.
|
|
| |
|
|
| |
SPECIFICATIONS: |
|
| |
|
|
| |
| Blower Flow Rate |
: |
More than 2.5 m3 / min free flow without resistance. |
| |
|
|
| Particulate Sampling |
: |
Less than 100u (RPM) with Size Selective Inlet (for model
HVF – 310) (Design approved by EPA & similar to Anderson
USA) |
| |
|
|
| Filter Holder |
: |
SS 304 to accommodate filter paper of 25 x 20 cm. |
| |
|
|
| Filter Media |
: |
Whatmann GF/A or EPM – 2000 |
| |
|
|
| Electronic Control Unit: |
: |
|
- Flow Rate
|
: |
From 0.25 M3 / Min on digital display |
| |
|
|
- Electronic Timer
|
: |
Sampling Time programmable upto 999 hrs. 59 mts. With Real
Time Clock |
| |
|
|
- Electronic Time Totalizer
|
: |
Upto 999 hrs. 59 mts. |
| |
|
|
- Volume Totalizer
|
: |
Upto 99999 M3 STP Corrected. |
| |
|
|
- Flow Measurement
|
: |
With exhaust orifice & Pressure sensor for flow |
| |
|
|
- Temp Measurement
|
: |
With Solid State Temperature Sensor for STP Correction. |
| |
|
|
- Display
|
: |
32 Character Alphanumeric LCD Module. |
| |
|
|
- Keys
|
: |
6 Nos. |
| |
|
|
- Status Indicator
|
: |
Under Voltage, Over Voltage & RUN LED’s |
| |
|
|
- Power Supply
|
: |
From 170 to 230 V AC, 50 Hz. |
| |
|
|
| Main Housing |
: |
Made up of sturdy Powder Coated aluminum. |
| |
|
|
| Brush Less Blower |
: |
Centrifugal regenerative Type Blower couple to 0.75 HP, 1
Ph AC Motor with 2” diameter Hose connection in a separate
weather proof housing. |
| |
|
|
| Gaseous Sampling |
|
|
| Flow Rate |
: |
0.3 to 3 LPM |
| Flow Control |
: |
Four way acrylic manifold with four inlet and one outlet and
four stainless steel needle valves for flow control of each
inlet. |
| Rotameter |
: |
0.3 LPM to 3 LPM with a resolution of 0.1 LPM. |
| Sampling Train |
: |
4 Nos. 35-ml glass impingers in a double – walled cold
box. |
| |
|
|
| Calibration |
: |
Each Blower & Gaseous Rotameter is individually calibrated
with Certified Calibrating Instruments with tracibility to NPL
New Delhi. Calibration Reports are provided. |
|
|
| |
|
|
| |
SPECIFICATION OF RESPIRABLE DUST
SAMPLER
PEM – HVF - 310 |
| |
|
|
| Blower Flow Rate |
: |
More than 2.5 m3 / min free flow without resistance. |
| |
|
|
| Particulate Sampling |
: |
Less than 10 ± 1 u with Size Selective Inlet (for RPM)
at 1.13 M3/Min
Less than 100 u with Gable Roof (for SPM). |
| |
|
|
| Filter Holder |
: |
SS 304 to accommodated filter paper of 25 x 20 cm. |
| |
|
|
| Filter Media |
: |
Whatman GF / A or EPM – 2000 |
| |
|
|
| Electronic Time Totalizer |
: |
Up to 99 Hrs. 59 Min. on LCD Module |
| |
|
|
| Keys |
: |
6 Nos. |
| |
|
|
| Constant Flow Controller |
: |
|
| Flow Rate |
: |
0.3 to 3 LPM |
| Flow Control |
: |
Four way acrylic manifold with four inlet and one outlet and
four stainless needle valve for flow control of each inlet. |
| Rotameter |
: |
0.3 LPM to 3 LPM with a resolution of 0.1 LPM |
| Sampling Train |
: |
4 Nos. 35-ml glass impingers in a double – walled cold
box. |
| |
|
|
| Calibration |
: |
Each Blower & Gaseous Rotameter is individually calibrated
with Certified Instruments. Calibration Reports are provided. |
|
|
| |
|
|
| |
| DIGITAL FLOW CALIBRATOR - Model PSI-DFC1 |
| |
| The Digital Flow Calibrator combines the simplicity &
versatility of a soap film bubble meter with the speed and accuracy
of a microprocessor. With handsfree operation, you can still
see the bubble, but you also have precision of a digital display
with accuracy of a quartz clock. No more tedious bubble formation,
bubble watching, timing & flow-rate/time calculation. The
DFC-1 accurately calibrates air sampling pumps (vacuum or pressure),
rotameters, orifices etc quickly & reliably with accurate
& reproducible results. |
| |
Features:
Primary
Standard
- Flow Accuracy ± 2%
- Wide Measuring Range
- Tracks Multiple Bubbles
- Auto – Ranging
- Single or Average Mode
- Run or Hold Mode
- Auto Shut-off
- Multi Function Self Diagnostic for Error free Operation
- Automatic Bubble-Feed for hand-free Operation
- Adjustable Glass Flow Cell for faster results at low flows
- Measures flow from pressure or Vacuum Source
- Field Replacable Glass-Flow Cell
- Fully Portable
- Mains or Battery Operated
- Optional Portable Printer for Permanent Records.
|
|
|
| |
|
|
| |
| CALIBRATORS FOR HI-VOL. SAMPLERS |
| |
| MODEL PEM - TLOC 1 |
| |
Polltech
make Calibrator Model PEM – TLOC 1 is traditional HVS
calibrator with a U Tube Manometer & offers a low cost
solution for calibration of HVS & RDS.
| Specifications : |
| Flow Range |
500 LPM to 2000 LPM |
| U tube Manometer: |
40 cms Glass Manometer on a Stand |
|
|
|
| |
|
|
| |
| MODEL PEM - TLOC 2 |
| |
Polltech make Calibrator Model PEM – TLOC 2 is a
state of art advanced calibrator with digital display of Flow
(STP corrected & uncorrected). It is based on the standard
design of EPA but uses Solid State Pressure & Temperature
sensor & advanced Micro controller to give digital
display of flow with an accuracy of better than 1%.
| Specifications : |
| Flow Range |
250 LPM to 2500 LPM |
| Display: |
16 x 2 alphanumeric LCD. |
| Keys: |
3 Nos. |
| Vacuum Sensor Range: |
0 to 1 PSI. |
| Temp. Range: |
5°C to 654°C |
|
|
|
| |
|
|
| |
|
|